Leak/flow calibration
and measurement
TM
Electronics
Topical Oxygen
Therapy

Virtual Leak Diameter (VLD) Flow Standards (cont)

Click here to download this document as a PDF

VLDs offer the technically correct solution to the control orcalibration of a gas or liquid flow. Ion medicare are specialist in fabricating small holes, calibrating and certifying, ready for use directly in a component or as fitting shown below.

Holes are 1micrometer and larger. Flow rates start at 0.01ml/min with maximum pressure 60bar. Holes are calculated and VLDs fabricated to specific customer applications. Tell us your testing specification and we will advise the VLD.

Once calibrated, the flow at any set controlled condition can be accurately predicted. Flow is accurate to within a fraction of 1%. Removable internal filtration protects the VLD. Applications include, Critical flow & pressure setting & control, particle separation, injection mixing, ink jet nozzles etc.

VLD DiscVLD Filter

VLD Disc

Assembly & Filter

VLD with quick fit maleVLD with quick fit male and femail

... with quick fit male

... with quick fit male and female

TECHNICAL SPECIFICATION

VLD:
Diameter:- 6mm
Thickness:- 1 to 1.7mm.
VLD sizes:- 1 to 1000micron.

FLOW RATE:-
0.01 to 1000ml/min, others on request.

MAXIMUM PRESSURE DIFFERENTIAL:-
60 bar.

FILTRATION:
(removable) 2.5 & 10micron particulate depending on VLD.

VACUUM & PRESSURE APPLICATIONS.

VLD MATERIAL:
Cu alloy.

HOUSING MATERIAL:
Stainless Steel.

THREAD SIZES:
1/8BSP male & Female.

'O' RING MATERIAL:
Viton